Copy this text
DEAFS – Deep-sea Autonomous and automated Fluid sampler data from the EMSO-Azores observatory, 2017
High temperature hydrothermal fluids are collected by the Deep-sea Autonomous and automated Fluid Sampler. The instrument is deployed at the footstep of the Montsegur edifice. The site, named Montserrat, expels hydrothermal fluid at 275°C which were collected beforehand by gas-tigh fluid sampler manipulated and triggered by the hydraulic arm of the submarine vehicle (ROV Victor6000).
Several chemical analyses are performed aboard the research vessel: such as salinity and d20/20 by refractometer, Conductivity, salinity and total dissolved solid (TDS) by electrode (standard solution at 1413 µS/cm), pH and Eh by electrode (pH standard 4.01 and 7; Eh standard solution at -220mV), H2S and total S concentrations measured by electrode (AquaMS, amperometric method) at measured pH value.
Regional Node: EMSO-Azores, Lucky Strike Hydrothermal Field
Vent field : Montserrat at the base of Montsegur
Latitude : 37º17.2744’N
Longitude : 32º16.5461’W
1699 m depth
Disciplines
Chemical oceanography, Environment
Keywords
Mid-Atlantic Ridge, EMSO-Açores, hydrothermal fluids, Lucky Strike, geochemistry, MOMARSAT 2017
Location
37.287907N, 37.287907S, -32.275768E, -32.275768W
Devices
Parameters
- high temperature hydrothermal fluid sampler
Location
- Regional node: Emso-Azores
- Site: Montserrat at Lucky Strike Hydrothermal Field
- Node: /
- Coordinates: N37º17.2744 and W32º16.5461’
- Depth: 1699m
Installation
- Cruise: SARRADIN Pierre-Marie, CANNAT Mathilde (2017) MOMARSAT2017 cruise, RV Pourquoi pas ?, https://doi.org/10.17600/17000500
- Dive: ROV 674 – 8 on July the 20th 2017
Recovery
- Cruise: CANNAT Mathilde (2018) MOMARSAT2018 cruise, RV L'Atalante, https://doi.org/10.17600/18000514
- Dive: ROV 698-3 on August the 12th 2018
Operating parameters
- Sampling period: every 27th of the month over a year time period
- Anti-fouling device : N/A
Sensor metadata
- Type: N/A
- Model: N/A
- Firmware: N/A
Sensor configuration
Sequenced fluid sampling programming
Test and calibration
- 12 Titanium bottles certified TOP Industrie S.A.S; Hydro test pressure @ 572bar; Working pressure : 400bar
Certification: N° 9615N, 9616N, 9617N, 9618N, 9619N, 9620N, 9621N, 9622N, 9623N, 9624N, 9625N, 9626N.
- Certification test of cylinder : working pressure : 412 bar; Ifremer PDG/IMN/SM/2IDM/17-0095
Data management
- Real time data transmission: No
Data
File | Size | Format | Processing | Access | |
---|---|---|---|---|---|
data | 872 octets | CSV | Raw data |